Instructions for Use
Electrosurgical Unit ME 102
14
V. 3.1
4.3
Mains Connection
For the assignment of the numbering, see section 4.2 “Functions of the Control Elements,
Connections and Displays”, page 12.
Before initial switching on of the unit, ensure that your grid complies with
the voltage setting of the unit indicated on the label (next to the power
input). If the voltages should not match, please contact our
Martin Service Center (see section 1.6 “Hotline”, page 8).
Potential adjustment to another supply voltage on site may only be carried
out by a person authorized to do so by Gebrüder Martin. Unauthorized
persons may not open the unit!
The unit is available in two versions:
•
100 – 127 VAC, T 2.0 AH
•
220 – 240 VAC, T 1.0 AH
The fuses can be found in the insert (17) on the mains connection socket.
Connect the unit to the power outlet using the mains cable. Set mains switch (1) on the front
side of the unit to position “ON”.
In order to be able to completely separate the unit from the grid in the event of danger, either
the power socket of the unit or the outlet into which the mains cable is plugged in should
remain accessible.
For the decommissioning of the unit, no special measures are necessary.
4.4
Equipotential Bonding Connection
For the assignment of the numbering, see section 4.2 “Functions of the Control Elements,
Connections and Displays”, page 12.
Equipotential bonding, aka potential equalization, is the conductive electrical
connection of the casings of units. It ensures that the units will always, even in
case of an electric defect, have the same potential. Equipotential bonding is
required for certain operating theaters, e. g. for intercardial operations, and can
be realized via the equipotential bonding connection (18). The necessary
connection cable is not included in the scope of delivery and can be purchased if
required.
4.5
Switching On / Off
For the assignment of the numbering, see section 4.2 “Functions of the Control Elements,
Connections and Displays”, page 12.
O
I
After switching on at the mains switch (1), the unit is operational. The surface of
the rocker switch must light up in green.