Instructions for Use
Electrosurgical Unit ME 102
20
V. 3.1
CAUTION
Risk of malfunction, risk of injury!
If the RF activation signal can be heard when no double foot switch or electrode handle is
connected, the unit is defective.
•
The unit must not be operated.
•
A technical inspection must be performed.
CAUTION
Risk of malfunction, risk of injury!
If the RF activation signal can be heard while the double switch or electrode handle is
connected but none of the control elements is triggered, one of these accessory parts is
defective.
•
This accessory must not be used.
•
It must be replaced.
4.7
Bipolar Operation
For the assignment of the numbering, see section 4.2 “Functions of the Control Elements,
Connections and Displays”, page 12.
4.7.1
Function of the Neutral Electrode
For bipolar application no neutral electrode is required. When selecting bipolar coagulation, the
monitoring of the neutral electrode is deactivated.
NOTICE
As a changeover from bipolar to monopolar application during the course of an operation is to
be expected, we recommend application of a neutral electrode under all circumstances.
4.7.2
Connecting Bipolar Instruments
All bipolar connection cables with small coax plug (e. g. for bipolar forceps, bipolar RF pliers)
can be connected to socket (14).