Instructions for Use
Electrosurgical Unit ME 102
16
V. 3.1
4.6.2
Connection of Handles
For monopolar cutting and coagulation, a handle with finger switch or a handle without a finger
switch can be connected together with a foot switch.
The handle
with finger switch
is connected to socket (3).
Handles with coax plug as well as handles with 3-pin plug can be
connected to socket (3).
•
Pay attention to the information in the accessories catalog of
KLS Martin.
Handles without finger switch or surgical RF instruments without
switch function (e. g. resectoscopes) are also connected to
socket (3). The corresponding double-pedal foot switch is
connected to socket (19) on the rear of the unit.
•
The desired active electrode must be inserted into the hexagonal guide of the handle.
—
The hexagon of the electrode snaps into place.
—
Rotation of the electrode during the application is prevented.
WARNING
Risk of electric shock!
When inserting the electrode and while replacing an electrode, the high-frequency current
must not be activated.