![Kitett MINIKIT SOLO Instructions For Use Manual Download Page 14](http://html1.mh-extra.com/html/kitett/minikit-solo/minikit-solo_instructions-for-use-manual_1973890014.webp)
29
GB
28
GB
13
Pictograms
0459
Compliance with Directive 93/42/EEC of 14 June 1993 relating to
medical devices.
Manufacturer.
Device reference.
Serial number of the device.
IP20
Degree of protection.
2: protected against penetration of foreign solid bodies with a
diameter > 12,5mm.
0: Not protected against water penetration.
Type BF equipment.
Continuous current.
Refer to the operating instructions
This symbol indicates the positioning of the batteries.
This symbol indicates the battery level.
Keep dry.
Fragile.
Store away from light.
This symbol indicates the temperature limit which must not be
exceeded in the case of operation, transport, and storage.
Symbols featured
Guidance and manufacturer’s declaration - electromagnetic immunity
Immunity test
Test level
IEC 60601
Compliance Level
Electrostatic discharge (ESD)
IEC 61000-4-2
±
8 kV through direct
contact
±
15 kV in the air
±
8 kV
±
15 kV
Fast transients and bursts
IEC 61000-4-4
±
2 kV for power lines
±
1 kV for input/output lines
±
2 kV
±
1 kV
Transient overvoltage
IEC 61000-4-5
±
1 kV between phases
±
2 kV between phase and
earth
±
1 kV between phases
±
2 kV between phase and
earth
Voltage dips, short interruptions
and voltage variations on power
supply input lines
IEC 61000-4-11
0% UT ; 0,5 cycle
At 0°, 45°, 90°, 135°, 180°,
225°,
270° and 315°
0% UT ; 0,5 cycle
At 0°, 45°, 90°, 135°, 180°,
225°,
270° and 315°
0% UT ; 1 cycle
ET 70% UT ; 25/30 cycles
Single phase : at 0°
0% UT ; 1 cycle
ET 70% UT ; 25/30 cycles
Single phase : at 0°
Magnetic field at power
frequencies (50/60 Hz)
IEC 61000-4-8
30 A/m
30 A/m
Radiated disturbances induced
by RF
IEC 61000-4-3
10 V/m
80 MHz to 2,7 GHz
10 V/m
80 MHz to 2,7 GHz
385 MHz at 27 V/m
450 MHz at 28 V/m
710, 745, 780 MHz at 9 V/m
810, 870, 930 MHz at 28 V/m
1720, 1845, 1970 MHz at
28 V/m
385 MHz at 27 V/m
450 MHz at 28 V/m
710, 745, 780 MHz at 9 V/m
810, 870, 930 MHz at 28 V/m
1720, 1845, 1970 MHz at
28 V/m
Conducted disturbances induced
by RF
IEC 61000-4-6
3V
0,15 MHz – 80 MHz
6V in ISM bands and bands
between 0.15 MHz and 80
MHz. 80% mA at 1kHz
3V
0,15 MHz – 80 MHz
6V in ISM bands and bands
between 0.15 MHz and 80
MHz. 80% mA at 1kHz