PERFORMANCE VERIFICATION TESTS
KI7000 UM-5
-
53 -
Model:
Instrument Connector:
APC [ ] PC [ ]
Report No.:
Date:
Fiber Type
:
Single mode [ ] Multimode [ ]
Range and Linearity Test
No
Attenuator setting
Power Meter Reading
Linearity Calculation
Acceptable Linearity Range
1
1
5.00 dB
-
2
15.00 dB
No 2 – No 1 =
9.50 ~ 10.50
3
20.00 / 25.00 dB
No 3 – No 2 =
4.50 ~ 5.50 / 9.50 ~ 10.50
4
35.00 dB
No 4 – No 3 =
9.50 ~ 10.50
Table 7: Test Record for Autotest Performance Verification Tests, KI7300C Series
Note1: The linearity range here is typically limited by attenuator linearity, rather than instrument performance.