PERFORMANCE VERIFICATION TESTS
KI7000 UM-5
-
42 -
All tests can be performed without access to the interior of the
instrument.
The test procedures described in this section are for performance
verification of a KI7303C InGaAs Loss Test Set.
Due to the large number of possible instrument configurations, it is
not possible to give detailed test procedures for all options in this
manual, so some parameters may need adjusting to the appropriate
specifications.
Required Equipment
This is the required equipment for the
performance test listed. Any equipment that satisfies the critical
specifications of the equipment given in the table may be substituted
for the recommended models.
Test Record
Results of the performance test may be tabulated on a
photocopy of the Test Record provided at the end of this test
procedure. It is recommended that you fill out the Test Record and
refer to it while doing the test. Alternatively, a soft copy of this
manual may be obtained from our web site.
Test Failure
If the equipment under test fails any performance test,
return the instrument to the nearest Sales/Service Office for repair.
Instrument Specification
Specifications are the performance
characteristics of the instrument that are certified, and are the limits
against which the equipment under test can be tested.
Any changes in the specifications due to manufacturing changes,
design or traceability to NATA will be covered in a manual change
supplement or revised manual. Such specifications supersede any
that were previously published.
General Instructions
Perform each step in the order given, using the corresponding test
equipment. Use Tables 1 ~ 3 to record general test details.
The SMF / MMF test lead fiber type and PC / APC connector polish
must be matched to the instrument type.
Make sure that all optical connections are dry and clean.
DO NOT
USE INDEX MATCHING OIL
. For cleaning, use the cleaning
instructions given in the section ‘Optical Connector’.
Make sure that all patch cords are fixed to the table so that they
won't move during measurements.
Ensure that the ambient conditions are in the following ranges:
Temperature:
21 ± 3 ºC
Relative humidity:
45 to 75 %
Summary of Contents for KI7300 Series
Page 1: ......
Page 57: ...QUICK REFERENCE GUIDE KI7600C Series Optical Power Meter KI7000 UM 5 55 ...
Page 59: ...QUICK REFERENCE GUIDE KI7400C 7800C Series Optical Light Source KI7000 UM 5 57 ...
Page 62: ...QUICK REFERENCE GUIDE KI7300C 7700C Series Loss Test Set KI7000 UM 5 60 ...
Page 64: ...NOTES KI7000 UM 5 62 ...
Page 65: ...NOTES KI7000 UM 5 63 ...
Page 66: ...NOTES KI7000 UM 5 64 ...