PERFORMANCE VERIFICATION TESTS
KI7000 UM-5
-
44 -
Light Source Section
Output Power (CW) Test
1.
Connect the equipment as shown in Figure 6.
Power Meter
LTS / Source
Figure 6. Test set-up for Output Power (CW) and Short Term Stability Tests
2.
Switch on the instruments.
3.
Set instruments to 1310 nm.
4.
Note the measured power level value in the test report in
Table 5
5.
Repeat the above on 1550 nm
Short Term Stability Test (optional)
1.
Connect the equipment as in Figure 6.
2.
Set instruments to 1310 nm.
3.
Let the light source to warm-up for 15 minutes, then note the
power.
4.
Record the optical power every 30 seconds for 3 minutes.
5.
Calculate max – min value (< 0.1 dB).
6.
Note this figure in Table 5
Alternatively, the instrument max/min recording function can be used
to record the data.
Time
Measured power, dBm
Drift, dB
Tick max / min values
00 sec
0.00
30 sec
60 sec
90 sec
120 sec
150 sec
180 sec
Summary of Contents for KI7300 Series
Page 1: ......
Page 57: ...QUICK REFERENCE GUIDE KI7600C Series Optical Power Meter KI7000 UM 5 55 ...
Page 59: ...QUICK REFERENCE GUIDE KI7400C 7800C Series Optical Light Source KI7000 UM 5 57 ...
Page 62: ...QUICK REFERENCE GUIDE KI7300C 7700C Series Loss Test Set KI7000 UM 5 60 ...
Page 64: ...NOTES KI7000 UM 5 62 ...
Page 65: ...NOTES KI7000 UM 5 63 ...
Page 66: ...NOTES KI7000 UM 5 64 ...