19
Compatibility EMC
The TD2 is intended for use in the electromagnetic environment specified below. The customer or
the user of the TD2 should assure that it is used in such an environment.
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete or ceramic tile.
If floor are covered with synthetic material,
the relative humidity should be at least 30%.
±2 kV for
power supply lines
±1 kV for
input /output lines
±2kV for
power supply lines
Mains power quality should be that of a typical
commercial or hospital environment.
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
±1 kV differential
mode
Mains power quality should be that of a typical
commercial or hospital environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment - guidance
Guidance and manufacture’s declaration – electromagnetic immunity
Electrical fast
transient /burst
IEC 61000-4-4
Surge
IEC 61000-4-5