EMC Table
Emission test
RF emission
Harmonic
Voltage
Group 1
Class B
Not applicable
Not applicable
The Digital Blood Pressure Monitor use RF energy only for its internal function. Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic equipment.
The Digital Blood Pressure Monitor is suitable for use in all establishments other than domestic and those directly connected to
the public low-voltage power supply network that supplies buildings used for domestic purposes.
CISPR 11
RF emissions
emissions
fluctuations/
flicker emissions
IEC 61000-3-3
IEC 61000-3-2
CISPR 11
Compliance
Electromagnetic environment – guidance
The Digital Blood Pressure Monitor is intended for use in the electromagnetic environment specified below. The customer of the user of the Digital Blood Pressure
Monitor should assure that it is used in such an environment.
Power frequency
magnetic field IEC 61000-4-8
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
3V/m
3V/m
(50Hz/60Hz)
Electrostatic
IEC 61000-4-2
discharge (ESD)
Immunity test
IEC 60601 test level
Compliance level
The Digital Blood Pressure Monitor is intended for use in the electromagnetic environment specified below. The customer or the user of Digital Blood Pressure Monitor
should assure that it is used in such an environment.
Guidance and manufacture’s declaration
Guidance and manufacture’s declaration – electromagnetic emission
NOTE UT is the a.c. mains voltage prior to application of the test level.
Floors should be wood, concrete or ceramic tile. If floor are covered with
synthetic material, the relative humidity should be at least 30%.
Power frequency magnetic fields should be at levels characteristic of a
typical location in a typical commercial or hospital environment.
Electromagnetic environment – guidance
The Digital Blood Pressure Monitor is intended for use in the electromagnetic environment specified below. The customer or the user of Digital Blood Pressure Monitor
should assure that it is used in such an environment.
Guidance and manufacture’s declaration – electromagnetic immunity
Guidance and manufacture’s declaration – electromagnetic immunity
17
Summary of Contents for BPX1A
Page 1: ...User Manual BPx1A...