4.5.
Configure the Modbus communication
4.5.1
Configure the slave number : F150
The F141 sub folder is displayed on the screen.
➢
Press Up key.
“F 150”
is displayed on the screen.
➢
Press
OK.
“F150”
flashes and the slave number is displayed below (the default number is 1).
➢
Press
OK.
The first digit of the slave number flashes.
➢
Press Up and Down keys to modify it then press
OK.
The second digit of the slave number flashes.
➢
Press Up and Down keys to modify it then press
OK.
The third digit of the slave number flashes.
➢
Press Up and Down keys to modify it then press
OK.
“F150”
flashes and the selected slave number is displayed below.
4.5.2
Configure the communication speed : F151
➢
Press Up and Down keys to go to the F151 sub folder.
The communication speed in bits per second is displayed (the default communication speed is 19200 bds).
➢
Press
OK.
The communication speed flashes.
➢
Press Up and Down keys and select the communication speed required between the following values :
•
2400 bds
•
4800 bds
•
9600 bds
•
19.2 Kbds
•
38.4 Kbds
•
115.2 Kbds
➢
Press
OK.
“F151”
flashes and the selected communication speed is displayed below.
4.6.
Activate the options
To activate an option, an activation code is necessary. This code is available nearby the company
distributing the device.
Available option for the ATE310 : high resolution in pressure (sub folder F170, only for the transmitters
having the option).
- High resolution in pressure option : F170
The sub folder “
F150”
is displayed on the screen.
➢
Press Up key.
“F170”
flashes on the screen and
“0”
is displayed below, meaning the option is not activated.
➢
Press
OK.
“0”
is flashing.
➢
Press up key.
The transmitter asks for an activation code.
➢
Enter the activation code (same process as the security code) then press
OK.
“F170”
flashes on the screen and “
1”
is displayed below, meaning the option is activated.
4.7.
Activate the digital input : F180
The ATE310 has a digital input. This digital input must be activated to enable the ATE310 to communicate via Modbus
(RS485) or Ethernet.
F 100
: Configure the transmitter
11