49
K
imberly
-c
larK
* Patient Warming Systems
M000 Control Unit Operator’s Manual: U .S .
Alarm/
Alert
Message
Displayed
Alarm/Alert
Description
System Status
Action
Primary
Outlet Water
Temperature
Sensor Shorted.
Alarm
T2 Thermistor
signal out of range
- low resistance.
Changes to Stop
Mode. Audio alarm
half-second tone
every 10 seconds.
Alarm/Alert
indicator flashing.
Contact Kimberly-Clark Service.
•
Secondary
Outlet Water
Temperature
Sensor Open.
Alarm
T1 Thermistor
signal out of range
- high resistance.
Changes to Stop
Mode. Audio alarm
half-second tone
every 10 seconds.
Alarm/Alert
indicator flashing.
Contact Kimberly-Clark Service.
•
Secondary
Outlet Water
Temperature
Sensor Shorted.
Alarm
T1 Thermistor
signal out of range
- low resistance.
Changes to Stop
Mode. Audio alarm
half-second tone
every 10 seconds.
Alarm/Alert
indicator flashing.
Contact Kimberly-Clark Service.
•
Inlet Water
Temperature
Sensor Open.
Alarm
T3 Thermistor
signal out of range
- high resistance.
Changes to Stop
Mode. Audio alarm
half-second tone
every 10 seconds.
Alarm/Alert
indicator flashing.
Contact Kimberly-Clark Service.
•
Inlet Water
Temperature
Sensor Shorted.
Alarm
T3 Thermistor
signal out of range
- low resistance.
Changes to Stop
Mode. Audio alarm
half-second tone
every 10 seconds.
Alarm/Alert
indicator flashing.
Contact Kimberly-Clark Service.
•
Outlet Water
Temperature
Sensor
T1 and T2 differ
by more than
1.0 degree C
Sensor disagreement can occur when unit
is heated from a starting temperature of less
than 20 °C. If this is the case, pre-heat the unit
to 25 °C & then continue to use per normal
operating instructions.
Contact Kimberly-Clark Service.
•
•
Outlet Water
Temperature
Overtemperature
Test Failure
(Control)
Alarm
The Control
Processor has
failed to detect a
simulated water
temperature fault.
Contact Kimberly-Clark Service.
•
Outlet Water
Temperature
Overtemperature
Test Failure
(Monitor)
Alarm
The Monitor
Processor has
failed to detect a
simulated water
temperature fault.
Contact Kimberly-Clark Service.
•