Alarm/
Alert
Message
Displayed
Alarm/Alert
Description
System Status
Action
Low bypass
flow. Heating
not available.
Alert
Insufficient
internal flow
during priming
or bypass mode.
Current Mode
continues. Audio
alarm half-second
tone every 2 min-
utes. Alarm/Alert
indicator flashing
Contact Kimberly-Clark Service to determine
cause of low bypass flow.
•
User settings
not saved.
System
default
settings
restored.
Alarm
Saved customer
user profile set-
tings are invalid.
Current Mode
continues. Audio
alarm half-second
tone every 2 min-
utes. Alarm/Alert
indicator flashing.
Check all custom user settings and temperature
alerts before restarting procedure. Save all values.
Contact Kimberly-Clark Service if problem persists.
•
•
User
Interface
Disconnect.
Alarm
User Interface
not detected by
control system.
Changes to Stop
Mode. Audio alarm
half-second tone
every 10 seconds.
Alarm/Alert
indicator flashing.
Check user interface cable connections on the
user interface module and K
imberly
-C
larK
*
Patient
Warming System M1000 Control Unit.
If the error persists, replace the User Interface
Cable and power up.
If problem continues, replace User Interface.
Contact Kimberly-Clark Service.
•
•
•
•
User
Interface
Disconnect.
Alert
User Interface
not detected by
control system.
Current Mode con-
tinues. Audio alarm
half-second tone
every 2 minutes.
Alarm/Alert
indicator flashing.
Check user interface cable connections on the
user interface module and K
imberly
-C
larK
*
Patient
Warming System M1000 Control Unit.
If the error persists, replace the User Interface
Cable and power up.
If problem continues, replace User Interface.
Contact Kimberly-Clark Service.
•
•
•
•
Patient
Temperature
Output
invalid.
Alarm.
Alarm In Manual
Mode, system is
unable to control
water within 1.0°C/
F of set point after
25 minutes in
current mode
or at the time
of last change
in Water Target
Temperature.
Changes to Stop
Mode. Audio alarm
half second tone
every 10 seconds.
Alarm/Alert
indicator flashing.
Contact Kimberly-Clark Service.
•
Patient
Temperature
Out system
error.
Alarm
Calculated Patient
Temperature
Output (Echo)
is ±0.1°C/F from
measured Patient
Temperature 1
(primary patient
temperature).
Changes to Stop
Mode. Audio alarm
half-second tone
every 10 seconds.
Alarm/Alert
indicator flashing.
Contact Kimberly-Clark Service.
•