M000 Control Unit Operator’s Manual: U .S .
. Remove the User Interface (Part # 0083-02) from its box . The clamp on the bot-
tom of the User Interface must be rotated 90 degrees before securing it to the
pole
(Figure 8 to Figure 9)
. Secure the User Interface to the cart pole .
2 . Connect the free end of the User Interface Cable to the plug at the base of the
User Interface and tighten the thumbscrews securely . Coil and secure cable
and place it on the holder just below the top of the tank .
3 . Remove the power cable (Part # 00833-0) from its bag and attach it to the
power outlet on the back of the CU (Figure 0) . Coil and secure the cable and
place it on the lower holder on the back of the CU .
4 . Remove the Fluid Delivery Line (Part # 00834-05) from its bag and attach it to
the back of the unit at the manifold . Insert the connector in the manifold (may
require jiggling the connector to seat it properly) then secure the connector by
moving the latch on the top of the manifold from right to left . (Figure ) .
5 . Remove the Patient Temperature Probe adapter (Part # 00835-02) from its bag
and attach the blue and grey connector to the Patient Temp Port on the back
of the unit . Patient Temp 2 Port is not used except for special situations where
dual patient temperature measurement is required . (Figure 3)
6 . Using the remaining butterfly nut and clamp, find the strain relief kit and
connect the port 2 cable to the second threaded bolt next to the power cable .
(Figure 2)
7 . Remove the Patient Temperature Output Cable [Part # 00835-52(Nellcor)] from its
bag and attach the black and grey connector to the Temp Out Port on the back
of the unit . (Figure 3)
i
nitial
t
est
. Turn the unit on by pressing the On/Off switch located above the power cable .
The unit will go through a self-check signaled by a single alarm . When the
self-check is complete, the unit will then check fluid level and an Alarm 4
(low water level) will sound . Silence the alarm by pressing the Alarm indicator
button . The unit will then display the Main Menu .
2 . Locate the container of cleaning solution from the cleaning kit
(Part # 00838-0) . Obtain an additional 2 liters of distilled or sterile water .
3 . On the User Interface, press the Down Arrow until the screen appears below .
Press Enter . The screen will then read, “Place the Fill Tube in the container of
water . Press Enter to fill .” Press Enter again .
Press the Down Arrow Key once until the screen reads, “Water Target
temperature XX, Press Enter to change .” Press Enter .
4 . Then take the end of the fill tube and place it in the cleaning solution . The
machine will aspirate the entire solution from the container . When the vial is
empty, transfer the fill tube to the 2 liters of distilled or sterile water obtained .
The unit will then aspirate the remaining fluid .