2-10
TOS
6200/6210
2
.7.
2
O
ptional
T
est
P
robe
LP01
-
TOS
/
LP02
-
TOS
•
N
ever connect the voltage measurement cable (thin wire) of the optional
LP01/LP02-
TOS
test probe to the
OUT
P
UT
terminals.
I
ts nominal sec-
tional area is inadequate for such currents, and burning may occur.
•
T
he rated current of the LP01-
TOS
and LP02-
TOS
test probes are up to
30
A
and 60
A
, respectively. Do not attempt to pass a current exceeding
the rated current.
•
I
mproper terminal connections can result in inaccurate measurements and
burns or injury resulting from heat generated by contact resistance at the
terminals.
1.
T
urn off the P
O
W
ER
switch of the
T
ester.
2.
R
emove the short-circuit bars connecting the
OUT
P
UT
and
SA
MPL
IN
G
terminals.
3.
C
onnect the L
O
W test lead to the L
O
W terminal of the
T
ester.
4.
C
onnect the each cable of the test probe to the
HI
G
H
and
RE
M
OTE
terminal of the
T
ester.
W
ARNIN
G
T
o
SA
MPL
IN
G
HI
G
H
terminal
Figure shows connections of the
TOS
6210 with the LP02-
TOS
.
T
o the
RE
M
OTE
terminal
T
o
OUT
P
UT
HI
G
H
terminal
T
o
SA
MPL
IN
G
L
O
W terminal
T
o
OUT
P
UT
L
O
W terminal
TOS
6210
Summary of Contents for tos6200
Page 1: ...Part No Z1 003 302 IB008755 Aug 2011 OPERATION MANUAL EARTH CONTINUITY TESTER TOS6200 TOS6210...
Page 36: ...3 8 TOS6200 6210...
Page 130: ...6 52 TOS6200 6210...
Page 136: ...7 6 TOS6200 6210...
Page 154: ...A 10 TOS6200 6210...