DME1500
SPECIFICATIONS
E-35
6.4
DC Current
6.5
True rms AC or AC+DC current
The burden voltage is the same as the DC current.
RANGE
RESOLUTION
ACCURACY
BURDEN VOLT-
AGE
500µA
0.01µA
0.2%+2
0.7Vmax.
5mA
0.1µA
0.2%+2
0.7Vmax.
50mA
1µA
0.2%+2
0.7Vmax.
500mA
10µA
0.2%+2
0.8Vmax.
2A
100µA
0.3%+2
0.8Vmax.
10A
1mA
0.3%+2
0.9Vmax.
Protection
500µA, 5mA, 50mA 500mA, 2A 5 ranges fuse protection.
10A range no fuse,15 seconds max.
Accuracy
Between 2% of range and full range.
RANGE
20Hz to
45Hz
45Hz to
2kHz
2kHz to
10kHz
10kHz to
20kHz
500µA
1%+15
0.5%+15
1%+15
2%+15
5mA
1%+15
0.5%+15
1%+15
2%+15
50mA
1%+15
0.5%+15
1%+15
2%+15
500mA
1%+15
0.5%+15
———
2A
1%+15
0.5%+15
———
10A
1%+15
0.5%+15
———
Protection
500µA, 5mA, 50mA, 500mA, 2A 5 ranges fuse
protection.
10A range no fuse,15 seconds max.
Crest Factor Range
3.0 at full scale
Summary of Contents for DME1500
Page 3: ...DME1500 J 1 1 000 V WARNING COUTION DANGER...
Page 4: ...J 2 DME1500 100V 200V Operation Manual Line Voltage...
Page 5: ...DME1500 J 3 D OFF 20kg G N L...
Page 6: ...J 4 DME1500 Check...
Page 10: ...J 8 DME1500...
Page 11: ...DME1500 J 9 1 1 1 1 1 2 1 3 AC 3P 2P 1 4 AC 0 08A T 200V 1 5 1...
Page 12: ...J 10 DME1500 1 2 0 50 40 70 6 11...
Page 13: ...DME1500 J 11 1 3 POWER OFF POWER ON...
Page 16: ...J 14 DME1500 1 6 D 2 1 3 2 3 2 3 2 2 GND...
Page 21: ...DME1500 J 19 3 3 1 dBm dBm 1mW dBm dBm 10 log10 1000 2 600 0 7746V 0dBm 3 2 DC 3 1 3 1...
Page 23: ...DME1500 J 21 3 3 AC DC rms 3 4 3 0 3 2 3 0 rms rms 2 2...
Page 24: ...J 22 DME1500 3 2 1 0 1 414 1 732 1 414 2 0 100 10 SCR 1 414 3 0 3 0 4 0 AC 3 0 9 0...
Page 27: ...DME1500 J 25 3 1 4 2 AC Hz dBm 5 V 10A MAX 10A 2A MAX 2A 6 2A 5 4...
Page 29: ...DME1500 J 27 5 5 1 POWER OFF 5 2 5 3...
Page 30: ...J 28 DME1500 5 4 2A 500 A 2A 1 POWER OFF 2 3 F2A 250V...
Page 42: ...E 6 Safety Symbols DME1500...
Page 46: ...E 10 INTRODUCTION DME1500...
Page 75: ......