Guidance and manufacturers declaration - electromagnetic immunity
The device in intended for use in the electromagnetic environment specifed below.
The customer or the user of the device should assure that it is used in such an environment.
Immunity test
IFC 60601 test level
Compliance level
Electromagnetic environment guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
+/- 6 kV contact
+/- 8 kV air
+/- 6 kV contact
+/- 8 kV air
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30%.
Electrical fast transient /
burst IEC 61000-4-4
+/- 2 kV for power supply
lines
+/- 1 kV for input / output
lines
+/- 2 kV for power supply
lines
Not applicable
Mains power quality should be that of a typical
commercial or hospital environment.
Surge IEC 61000-4-5
+/- 1 kV line(s) to line(s)
+/- 2 kV line(s) to earth
+/- 1 kV differential mode
Not applicable
Mains power quality should be that of a typical
commercial or hospital environment.
Voltage dips, short interruptions
and voltage
variations on power supply input
lines
IEC 61000-4-11
<5% UT (>95% dip in UT)
0,5 cycle
40% UT (60% dip in UT)
5 cycles
70% UT (30% dip in UT)
25 cycles
<5% UT (>95% dip in UT)
5 s
<5% UT (>95% dip in UT)
0,5 cycle
40% UT (60% dip in UT) 5
cycles
70% UT (30% dip in UT) 25
cycles
<5% UT (>95% dip in UT)
5 s
Mains power quality should be that of a typical
commercial or hospital environment. If the user
of the device requires continued operation during
power mains interruptions, it is recomended that
the device be powered from an uninterruptable
power supply or
a battery.
Power frequency (50/60 Hz)
magnetic feld
IEC 61000-4-8
3 A/m
3 A/m
The power frequency magnetic felds should be
at levels characteristic of a typical commercial or
hospital environment.
NOTE UT is the a. c. mains voltage prior to application of the test level.