USE OF THE PRODUCT
Mattress is suitable for persons up to 110 kg body weight. If overloaded, mattress do not work properly or damage to mattress or user
may occur. Air pressure in the mattress should be adjusted depending on the weight of user. We suggest to use table below in order
to adjust proper air pressure in the mattress.
User‘s weight
Position of air pressure knob
<40 kg
2-3
40 - 60 kg
3-4
60 - 70 kg
4-5
70 - 80 kg
5-6
80 - 90 kg
6-7
90 - 110 kg
8
WARNING! In case the pressure is constantly low, check for any holes (on the interchangeable elements or on the
connection tubes). If it is necessary substitute the damaged connection tubes or the interchangeable elements, otherwise
contact your dealer! If a rapid deflation is necessary unplug the connection tube from the mattress!
WARNING! IF MATTRESS IS NOT BEING USED FOR A LONG PERIOD, PLEASE PREPARE IT FOR STORAGE
• Switch off power from main and unplug, separate the pump with mattress and put the mattress flat.
• Fold the mattress in two from mattress foot to head
• Allow time for air to escape from mattress, and then reduce the space to storage
CLEANING AND DISINFECTION
Clean the mattress with mild soap. Do not immerse or soak pump unit. After cleaning, dry the mattress without direct sunlight
exposure. Check the connection tubes are not obstructed by dust or else. In this case use a soft swab to clean the tubes.
WARNING! Turn pump off before starting cleaning procedure
WARNING! Never use acids, alkalis or solvents such as acetone