24
Keysight X8712A IoT Device Battery Life Optimization Solution User’s Guide
N6705C Four-Channel Configuration
This setup is recommended for multi-event synchronization testing for both
signaling and normal operation. Normal operation is defined as the operation
upon successful connection.
This four-channel configuration requires the following setup:
– Channel 1: Connect DUT
– Channel 2: Connect X8712AD RF event detector
– Channel 3 and 4: Custom wiring and/or custom sensors with voltage or
current response to synchronize the events
Channel 1 is fixed as 'Battery Emulator' module and is used to power up the DUT
and monitor the total current drain.
Channel 2 is fixed as 'Power Detector' module with the X8712AD RF event
detector connected to capture the RF signal transmitted or received by the DUT.
1
Follow the instructions from
of the
DUT in Self Advertising or Test Mode
section.