Language Reference
5
Series N8700 User’s Guide
75
Source Commands
Source commands program the voltage, current, triggered, and
protection functions.
[SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude] <value>|MIN|MAX
[SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude]? [MIN|MAX]
[SOURce:]CURRent[:LEVel]:TRIGgered[:AMPLitude] <value>|MIN|MAX
[SOURce:]CURRent[:LEVel]:TRIGgered[:AMPLitude]? [MIN|MAX]
These commands set the immediate and the triggered output current.
Values are programmed in amperes. The immediate level is the
output current setting. The triggered level is a stored value that
transfers to the output when a trigger occurs. The *RST values = Min.
Model ( I rating)
400A
330A
220A
165A
110A
85A
55A
42A
33A
22A
11A
5.5A
Minimum current
0
0
0
0
0
0
0
0
0
0
0
0
Maximum current
420
346.5
231
173.25
115.5
89.25
57.75
44.1
34.65
23.1
11.55
5.775
Model ( I rating)
250A
170A
125A
85A
65A
50A
34A
17A
8.5A
Minimum current
0
0
0
0
0
0
0
0
0
Maximum current
262.5
178.5
131.25
89.25
68.25
52.5
35.7
17.85
8.925
[SOURce:]CURRent:PROTection:STATe ON|OFF
[SOURce:]CURRent:PROTection:STATe?
This command enables or disables the over-current protection (OCP)
function. The enabled state is On (1); the disabled state is Off (0). If
the over-current protection function is enabled and the output goes
into constant current operation, the output is disabled and OC is set
in the Questionable Condition status register. The *RST value = Off.
An over-current condition can be cleared with the Output Protection
Clear command after the cause of the condition is removed.
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude] <value>|MIN|MAX
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]? [MIN|MAX]
[SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude] <value>|MIN|MAX
[SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude]? [MIN|MAX]
These commands set the immediate and the triggered output voltage.
Values are programmed in volts. The immediate level is the output
voltage setting. The triggered level is a stored value that transfers to
the output when a trigger occurs. The *RST values = Min.
The range of values that can be programmed for these commands is
coupled with the voltage protection and the voltage limit low settings.
The maximum value for the immediate and triggered voltage level is
either the value in the following table, or the voltage protection
setting divided by 1.05; whichever is lower. The minimum value is
either the value in the table, or the low voltage setting divided by
0.95; whichever is higher.