Calibrations Procedures

 

Transfer function calibrations

Purpose and Procedure

The signal path for the Type C Tx tests always includes additional cables 

and switch relays, that differ from the defined setup of the pure standard 

tests. Because of this, additional calibrations are required.

The 

Check Transfer function 

procedures are meant to verify the right 

location of the transfer functions. They don't calculate or measure any 

S-parameter or transfer functions.

There is a separated calibration for each signal path as shown in 

Figure 16

Figure 16

Transfer function calibrations

Parameters

Most calibration and test procedures as well as the groups containing 

them, have parameters that control the details of how the procedures are 

run. First, select a specific calibration or test procedure or one of the 

groups containing in the N5990A procedure tree. The parameters are 

displayed in a property list on the right side of the screen. If they are not 

displayed, click the 

Properties

 button. Depending on the user selection on 

the left side of the top of the list, the list is either ordered alphabetically or 

in categories.

Summary of Contents for N5990A Type-C

Page 1: ...User Guide Keysight N5990A Type C Multi Standard Solution...

Page 2: ...quirements beyond those set forth in the EULA shall apply except to the extent that those terms rights or licenses are explicitly required from all providers of commercial computer software pursuant t...

Page 3: ...ion N5990A Type C Setup 12 Remote Instruments Server 13 SigTest for USB 14 Driver for DPR 100 15 3 N5990A Type C Station Configuration N5990A Type C Station Configuration 18 Type C Station Selection 1...

Page 4: ...dard combination USB RX USB TX 39 Standard combination USB RX USB TX DP T 39 6 Test Procedures PDO testings 42 Parameters 42 Connection Diagrams 44 Standard combination USB TX USB RX 44 Standard combi...

Page 5: ...m 6 Requirements 7 Related documents 8 Used Open Source Software 9 This guide provides a detailed description of the N5990A Test Automation Software Platform The BitifEye ValiFrame Test Automation sof...

Page 6: ...to use the ValiFrame Type C Multi Standard Solution The ValiFrame Type C Multi Standard Solution runs the tests of several Type C Alt mode standards in one run The software supports automatic control...

Page 7: ...Platform Support N5990A 004 Type C Alt Mode Receiver Tests and PDO Support N5990A 006 Multi Standard Sequencing N5990A 007 Test Sequencer N990A 010 USB Receiver Tests N5990A 102 Interface to USB Trans...

Page 8: ...ntroduction Related documents For detailed description of the test with the included standards USB DP refer to the following documents Keysight N5990A Test Automation Software Platform for USB Keysigh...

Page 9: ...ntlr org BSD licence Ionic ZipLib https github com jstedfast Ionic Zlib Apache license ZedGraph https sourceforge net projects zedgraph LGPL ZipLib https icsharpcode github io SharpZipLib GNU General...

Page 10: ...10 Keysight N5990A Type C Multi Standard Solution User Guide 1 Introduction...

Page 11: ...Keysight N5990A Type C Multi Standard Solution User Guide 2 Installation N5990A Type C Setup 12 Remote Instruments Server 13 SigTest for USB 14 Driver for DPR 100 15...

Page 12: ...90A Software installer file has been executed a window wizard appears see Figure 1 It is necessary that all the software which are required to run ValiFrame must be installed before accepting the lice...

Page 13: ...Type C controller N7018A and to have all necessary USB driver binaries It must be installed at the PC where the N7018A s control USB cable is pluged in If necessary download it from http www bitifeye...

Page 14: ...stallation SigTest for USB The SigTest software is used for USB tests and calibrations It must be installed at the PC where ValiFrame is running as well as at the scope where the USB TX app is running...

Page 15: ...tall a driver This driver must be installed in the same oscilloscope where the DisplayPort transmitter App is installed If necessary download it from https www unigraf fi support downloads software do...

Page 16: ...16 Keysight N5990A Type C Multi Standard Solution User Guide 2 Installation...

Page 17: ...ion 18 Type C Station Configuration 21 Instruments Configuration 24 Refer to the N5990A Getting Started Guide for instructions on how to install and start the ValiFrame Test Automation software platfo...

Page 18: ...e Type C and the set of instruments used for it To start the Station configuration double click of the short cut icon see Figure 3 which can be found on the Desktop Alternatively start the application...

Page 19: ...ntries displayed in the Application Server with the actual IP address and port number of the database application server Results Viewer Select either Excel or HTML to view the results after the tests...

Page 20: ...ound option that appears in the drop down list None deactivates the Play button the application does not play any audio for the corresponding action Car brake Feep Feep Ringing TaDa Tut After setting...

Page 21: ...e C Controller and Power Supply from the drop down options Figure 5 Type C Station Configuration Combined Standards The ValiFrame Type C Module can support two different standard combinations USB RX U...

Page 22: ...performed with two different kinds of power supply N6786A The N6786A is suitable for testing sink and source up to 20V 3A 5V power supply The simple 5V power supply is just to bring a DUT to work but...

Page 23: ...ending on the selected power supply for example see Figure 6 Figure 6 USB Station Configuration for Type C For more detailed description of the parameters of each standard refer to the following relat...

Page 24: ...e 7 Type C Instrument Configuration In the example of the Figure 7 ValiFrame is running at the controller PC of the M8041A Also the Type C controller N7018A is connected to this PC so the Remote Instr...

Page 25: ...lt in Offline mode To change the state to Online select the corresponding check box of instrument If you click Check Connections all selected addresses will be queried Because it s not possible to run...

Page 26: ...26 Keysight N5990A Type C Multi Standard Solution User Guide 3 N5990A Type C Station Configuration...

Page 27: ...Keysight N5990A Type C Multi Standard Solution User Guide 4 N5990A DUT Configuration Starting the N5990A Type C 28 PDO Configuration 30 Configuration of the Sub stations 33...

Page 28: ...A Protocol Name double click the N5990A Type C icon that appears on the desktop as shown in Figure 8 Alternatively start the N5990A Type C station from Start All Programs BitifEye Type C ValiFrame Typ...

Page 29: ...e C Multi Standard Solution User Guide 29 N5990A DUT Configuration 4 From the main windows click the Configure DUT button The Type C configure DUT dialog will appear as shown in Figure 10 Figure 10 Ty...

Page 30: ...nsumer PDOs are supported by a fixed 5V supply The unsupported PDOs are displayed but they cannot be selected see Figure 12 Figure 12 Supported PDOs Select the PDOs you want to include intPower Supply...

Page 31: ...ity However you can also type the actual current capability of the power supply This 5V power supply may be a high power USB plug in charger If the N67xx is used the dialog shows the available channel...

Page 32: ...e channel to CC1 for normal connection orientation and to CC2 for flipped orientation If you want to test with flipped SBU lines select the corresponding check box of the Flip SBU lines option The SBU...

Page 33: ...ialogs some parameters are predefined for the Type C tests and cannot be changed However some parameters are editable also and can be modified as per the need For more detailed description of the para...

Page 34: ...34 Keysight N5990A Type C Multi Standard Solution User Guide 4 N5990A DUT Configuration...

Page 35: ...Keysight N5990A Type C Multi Standard Solution User Guide 5 Calibrations Procedures Transfer function calibrations 36 Calibrations for Rx Tests 39...

Page 36: ...S parameter or transfer functions There is a separated calibration for each signal path as shown in Figure 16 Figure 16 Transfer function calibrations Parameters Most calibration and test procedures a...

Page 37: ...lowing are the parameters FilePathTransferFucntion It is the name of the transfer function file which is used in the affected calibration or test procedure You can select this filename by clicking the...

Page 38: ...one specific setup You can copy those transfer functions from your ValiFrame PC to the scope Figure 19 Default Transfer Functions But you must be aware that your own setup may have different properti...

Page 39: ...ion USB RX USB TX see Station Configuration selection Figure 5 doesn t include any switch relays and the calibrations are made at the same point as for the pure USB Rx calibrations Therefore the calib...

Page 40: ...40 Keysight N5990A Type C Multi Standard Solution User Guide 5 Calibrations Procedures Figure 21 Connection diagram for calibrations of the standard combination USB RX USB TX DP TX...

Page 41: ...iFrame USB configured for USB3 1 with a Keysight Type C fixture N7015A The DP test procedures are also configured in the same way as they are in pure DP Tx tests ValiFrame Type C not only configures t...

Page 42: ...rameters that control the details of how the procedures are run First select a specific calibration or test procedure or one of the groups containing in the N5990A procedure tree The parameters are di...

Page 43: ...23 PDO test iteration Save a Workbook after each PDO step If it is set to true for each single PDO the set of test results is saved in an html workbook Path for autosaved Workbooks Set here the locati...

Page 44: ...for each test depending e g if currently TX or RX test is running This leads to complex setups which differ depending on the standard applications you want to integrate Standard combination USB TX US...

Page 45: ...xture data inputs Connect the Keysight N7015A Fixture to the Keysight N7018A Type C Test Controller Connect the Keysight N7015A Fixture data outputs to the relays of the switch modul in slot 4 of the...

Page 46: ...er power input Connect the Keysight N7018A Type C controller to the ValiFrame PC or a PC where a Remote Instruments Server is running Don t connect the N7018A to the scope Use blocking capacitors Keys...

Page 47: ...host at both lanes in one run Once the calibrations are done there s no reconnection needed during the test run Because the USB RX signal path is different from the path at a standard USB RX test we h...

Page 48: ...R 100 is connected to the scope The Type C controller N7018A must not be connected to the scope It is controlled by the ValiFrame PC or a PC where the Remote Instruments Server is running It is recomm...

Page 49: ...mbination USB RX USB TX DP TX 55 The USB TX tests with the Keysight Type C fixture N7015A require the use of transfer functions To get comparable USB RX test results with the USB IF fixture left and t...

Page 50: ...ion of the used transfer function for the relevant test cases Infiniisim feature can be used with Keysight scope to calculate the transfer function For more details refer to the documentation of Infin...

Page 51: ...Appendix 7 For best test results you should measure the S parameters for specific setup must be measured with a Network Analyzer or with the precision cable feature of the Keysight scope For details a...

Page 52: ...p5dbL ossAt2p5GHz_Embed tf4 10G Device C C SSGen2_TxComp12p2dB_Embed ding s4p N7015A_WithPlug_PP S4P N7015A_Deembed__TypC_Typ C_14p1dbAt5GHz_Embed tf4 Captive Device Mock_Device_Cascaded_Model_8 p5dB_...

Page 53: ...P S4P BIT2100A_1Stage s4p N7015A_Deembed__13p5dbL ossAt2p5GHz_Embed__1Stag eSwitch_Deembed tf4 embed Channel S Parameter channel embed S Parameter fixture de embed S Parameter Switch de embed Transfer...

Page 54: ...ard embed Channel S Parameter channel embed S Parameter fixture de embed S Parameter Switch de embed Transfer Function C C SSGen2_TxComp12p2dB_Embed ding s4p N7015A_WithPlug_PP S4P BIT2100A_1Stage s4p...

Page 55: ...the more complex setup for the standard combination USB RX USB TX DP TX two or three switch stages must be taken into account depending on the tested standard USB TX or DP TX and lane For the DP test...

Page 56: ...56 Keysight N5990A Type C Multi Standard Solution User Guide 7 Appendix...

Page 57: ...Keysight N5990A Type C Multi Standard Solution User Guide 57...

Page 58: ...This information is subject to change without notice Keysight Technologies 2018 Edition1 0 April 2018 www keysight com...

Reviews: