Step 5: Verify Operation of the Module
To Perform Self Test
You can perform the self test by using the M960x PXIe Source Measure Unit Soft
Front Panel (the M960x SFP).
NOTE
Before performing the self test, turn the instrument output off and disconnect
cables from the measurement terminals.
1. Click Start menu > Keysight M960x Source Measure Unit > M960x SFP. The
“Connect to Instrument” window will open.
2. In the list on the window, highlight the M9601A modules to connect and click
the “Connect” button to launch the M960x SFP.
3. Confirm that the Access indicator on the M9601A front panel turns green.
Refer to
and
4. Select “Self Test...” from the Utilities menu. The “Self Test” window will open.
5. Click the “Run Self Test” button at the bottom of the “Self Test” window to
start the self test. When the self test is completed, the results will be
displayed in the “Self Test” window.
NOTE
If the module fails the self test, it will be defective. See
to return the modules to Keysight.
CAUTION
Use the module within the rating
If you input a signal above the rating, the module will be damaged.