Keysight M9601A Startup Guide, Edition 2
19
Front Panel Features
LED Indicators
The M9601A module has two LED indicators, Access and Status. These
indicators show the status of the M9601A. See
for the meanings. If the
emergency shutdown occurs, both indicators turn red.
Table 1-3
LED Status
Measurement Terminals
The M9601A has the following measurement terminals. You can make 2-wire
connection using the High force and Low force terminals or 4-wire connection
(Kelvin connection) using the High force, High sense, Low force and Low sense
terminals. The Kelvin connection is effective for high current measurement.
WARNING
There are potentially hazardous voltages (
210 V) present at the High force, High
sense, and Guard terminals of this instrument. To prevent electrical shock, the
following safety precautions must be observed during the use of the instrument.
LED
LED State
Status
Access
Solid Green
The module is turned on, the initialization is
completed, and the operation is ready.
Flashing Green
The “Identify PXIe module” feature is on.
Solid Red
Emergency shutdown occurs.
Off
The module is turned off.
Status
Solid Green
Output is enabled and the output voltage is within
42 V.
Solid Yellow
Output is enabled and the output voltage is over
42 V.
Solid Orange
Output is enabled and the output voltage or
current reaches the compliance.
Solid Red
Emergency shutdown occurs.
Off
Output is disabled.