Service
5. Remove the side shield from the defective module.
a. Using a Philips #1 (PH1) screwdriver, remove the two screws that secure the side shield to the
module.
NOTE: Keep the screws
; extra screws are not included with the replacement module.
b. Slide the shield toward the module's front panel. This aligns the engagement tabs so you can remove
the shield.
38
Agilent M9360A PXI Attenuator/Preselector Service Guide