Performance Verification Tests
Test Procedure
Test Setup
1. Turn on the VNA and allow it to warm up for 90 minutes.
2. Install the M9360A into the PXI chassis.
3. Turn on the PXI chassis and allow it to warm up for 15 minutes.
4. Connect the cables and adaptors to the VNA so that the ends of both cables have 3.5 mm male connectors.
5. Preset the VNA.
6. Configure the VNA as shown in
Table 3
.
Table 3
Parameter
Value
Notes
Start Frequency
10 MHz
Stop Frequency
26.5 GHz
RF Output Power
-15 dBm
Use preset value
IF Bandwidth
1 kHz
Number of Trace Points
601
Calibrate the Vector Network Analyzer
There are two options for calibrating the VNA: use Option A or Option B as noted below.
Option A
. Using an Agilent N4691B ECal module:
1. Connect the ECal module to the VNA. ECal port A connects to VNA PORT 1, and ECal port
B connects to VNA PORT 2. Connect the USB cable between the ECal module and the
VNA.
2. Allow the ECal to warm up and indicate a READY status.
3. Perform a full two-port calibration on the VNA.
4. Press the
Cal
key on the VNA.
5. From the softkey menu, press
Start Cal
, and then
Cal Wizard
.
6. In the
Calibration Wizard Begin
dialog box, select
Use Electronic Cal
(ECal) and then
click the
Next
button.
7. Follow the ECal wizard instructions for a full two-port calibration, relying on the calibration
data stored in the ECal module.
Option B
. Using an Agilent 85052B Calibration Kit:
1. In the
Calibration Wizard Begin
dialog box, click
SmartCal
(GUIDED Calibration), use
Mechanical Standards
, and then click the
Next
button.
2. Follow the wizard instructions for a full two-port calibration.
3. When prompted for the type of thru calibration, select
Unknown Thru Cal
.
4. Set the VNA set
Single Trigger
mode.
5. Set the VNA Meas to S21 (measurement mode).
6. Set the VNA Format to
Log Mag
.
Agilent M9360A PXI Attenuator/Preselector Service Guide
25