Performance Verification Tests
5. Set the RF power of the signal source to ON.
6. Configure the M9360A as shown in
Table 5
.
Table 5
Parameter
Value
Preselector Enabled
No
RF/LO Out
1 (Low band)
Step Attenuator
0 dB
7. For each frequency listed in
Table 6- Test Points:
Table 6 - Test Points
Frequency
250 kHz
1 MHz
2 MHz
5 MHz
9.9 MHz
a. Set the power meter frequency. From the N1914A front panel, press the Channel button, highlight the
Frequency field, and enter the frequency for both channels.
b. Note the displayed power for both channels.
c. Calculate the insertion loss as: Channel A power – Channel B power.
d. Record the insertion loss value for RF 1 OUT in the Test Record Card (see page
).
RF IN to RF 2 OUT measurement
1. Move the Channel B power sensor to the M9360A RF 2 OUT port, using the right angle SMA and 3.5 mm
(m) to 3.5mm (m) adaptors.
2. Using the M9360A Soft Front Panel, enable
Custom Settings
and set the M9360A
RF/LO Out
to 2 (High
band).
3. Repeat steps 7.a. to 7.c. from
RF IN to RF 1 OUT measurement
in the above procedure.
4. Record the insertion loss value for RF 2 OUT in the Test Record Card (see page
).
LO IN to LO 1 OUT measurement
1. Move the Channel A power sensor to the LO IN port on the M9360A.
2. Move the Channel B power sensor to the LO 1 OUT port on the M9360A, using the right angle SMA and
3.5 mm (m) to 3.5mm (m) adaptors.
3. Using the M9360A Soft Front Panel, set the M9360A
RF/LO OUT
to 1 (Low band).
4. Repeat steps 7.a. to 7.c. from
RF IN to RF 1 OUT measurement
in the above procedure.
5. Record the insertion loss value for LO 1 OUT in the Test Record Card (see page
).
LO IN to LO 2 OUT measurement
1. Move the Channel B power sensor to the LO 2 OUT port on the M9360A, using the right angle SMA and
3.5 mm (m) to 3.5mm (m) adaptors.
Agilent M9360A PXI Attenuator/Preselector Service Guide
31