Performance Verification Tests
Test Configuration
Figure 1
Test Procedure
Test Setup
1. Turn on the signal source and allow for a warm up of 15 minutes.
2. Install the M9360A into the PXI chassis.
3. Turn on the PXI chassis and allow it to warm up for 15 minutes.
4. Preset all equipment.
5. Connect the equipment as seen in
Figure 1
.
a. Connect the signal analyzer 10 MHz OUT port to the signal source 10 MHz IN port.
b. Connect the RF OUT of the signal source to the RF IN port of the M9360A preselector.
c. Connect the spectrum analyzer to the RF 2 OUT port of the M9360A preselector.
6. Open the M9360A Soft Front Panel and configure as shown in
Table 3
.
Table 3
Parameter
Value
Center (Input) Frequency
See
Table 6 - Test Points
Input Level
-40 dBm
Custom Settings
Unchecked
Preselector Enabled
Checked
7. Configure the signal source as shown in .
Table 4
.
Table 4
Parameter
Value
Start Frequency
M9360A Center Frequency - 0.1 GHz
Stop Frequency
M9360A Center Fre 0.1 GHz
Dwell Time
10 ms
Points
1000
8. Configure the signal analyzer as shown in
Table 5
.
Agilent M9360A PXI Attenuator/Preselector Service Guide
33