Performance Verification Tests
2. Configure the M9360A Soft Front Panel as shown in
Table 6
.
Table 6
Parameter
Value
Notes
RF/LO OUT
1 (Low-band)
This setting applies to the RF 1 OUT path.
3. Configure the VNA as shown in
Table 7
.
Table 7
Parameter
Value
Notes
Stop Frequency
2.9 GHz
According to the specification of RF 1 OUT.
Measure RF 1 OUT Insertion Loss ≤ 2.9 GHz
1. Take a sweep on the VNA.
2. Select
Scale
>
AutoScale
on the VNA.
3. Select
Marker Search
>
Min
(perform min marker search).
4. Record the marker value as the Measured Insertion Loss (≤ 2.9 GHz) of the RF 1 OUT path (use the Test
Report Card – page
).
Measure RF 1 OUT Insertion Loss 2.9 GHz to 26.5 GHz
1. Configure the VNA as shown in
Table 8
.
Table 8
Parameter
Value
Notes
Start Frequency
2.9 GHz
According to the specification of RF 1 OUT.
Stop Frequency
26.5 GHz
According to the specification of RF 1 OUT.
2. Take a sweep on the VNA.
3. Select
Scale
>
AutoScale
on the VNA.
4. Select
Marker Search
>
Min
(perform min marker search).
5. Record marker value as the Measured Insertion Loss (> 2.9 GHz) of the RF 1 OUT path (use the Test
Report Card – page
).
Measure RF 2 OUT (bypass) Insertion Loss
1. Disconnect the cable from RF 1 OUT on the M9360A, and connect it to RF 2 OUT on the M9360A.
2. Configure the M9360A Soft Front Panel as shown in
Table 9
.
Table 9
Parameter
Value
Notes
RF/LO OUT
2 (High-band)
This setting applies to the RF 2 OUT path.
3. Configure the VNA as shown in
Table 10
.
Table 10
Parameter
Value
Notes
Start Frequency
10 MHz
According to the specification of RF 2 OUT(bypass).
Stop Frequency
26.5 GHz
According to the specification of RF 2 OUT(bypass).