Testing Performance
2
DSOX1204A/G Oscilloscopes Service Guide
23
The difference in average voltage readings should be within the test limits of
.
If a result is not within the test limits, go to the “Troubleshooting” chapter. Then
return here.
11
Disconnect the oscilloscope calibrator from the oscilloscope.
12
Repeat this procedure to check the DC vertical gain accuracy with the
remaining Volts/div setting values in
.
13
Finally, repeat this procedure for the remaining channels to be tested.
Use a Blocking Capacitor to Reduce Noise
On the more sensitive ranges, such as 0.5 mV/div, 1 mV/div, 2 mV/div, and
5 mV/div, noise may be a factor. To eliminate the noise, add a BNC Tee, blocking
capacitor, and shorting cap at the oscilloscope channel input to shunt the noise to
ground. See
. If a BNC capacitor is not available, use an SMA blocking
capacitor, adapter, and cap. See
“Blocking capacitor and shorting cap
equipment list on
for details.
Figure 2
Using a Blocking Capacitor to Reduce Noise
To oscilloscope input
BNC shorting
cap
Blocking
Capacitor
Summary of Contents for InfiniiVision DSOX1204A
Page 1: ...Keysight InfiniiVision DSOX1204A G Oscilloscopes Service Guide ...
Page 8: ...8 DSOX1204A G Oscilloscopes Service Guide Figures ...
Page 10: ...10 DSOX1204A G Oscilloscopes Service Guide Tables ...
Page 36: ...36 DSOX1204A G Oscilloscopes Service Guide 2 Testing Performance ...
Page 48: ...48 DSOX1204A G Oscilloscopes Service Guide 5 Replacing Assemblies ...
Page 50: ...50 DSOX1204A G Oscilloscopes Service Guide 6 Replaceable Parts ...
Page 56: ...56 DSOX1204A G Oscilloscopes Service Guide Index ...