22
DSOX1204A/G Oscilloscopes Service Guide
2
Testing Performance
Figure 1
Setting up Equipment for DC Vertical Gain Accuracy Test
7
Adjust the output so that the multimeter reading displays, as close as possible,
the first Volts/div calibrator setting value in
8
Wait until the measurement settles.
9
Read the “current” average voltage value again as V2.
10
Calculate the difference V2 - V1.
Oscilloscope
Oscilloscope
Calibrator
Digital
Multimeter
BNC Tee
BNC (f) to dual
bananna adapter
Summary of Contents for InfiniiVision DSOX1204A
Page 1: ...Keysight InfiniiVision DSOX1204A G Oscilloscopes Service Guide ...
Page 8: ...8 DSOX1204A G Oscilloscopes Service Guide Figures ...
Page 10: ...10 DSOX1204A G Oscilloscopes Service Guide Tables ...
Page 36: ...36 DSOX1204A G Oscilloscopes Service Guide 2 Testing Performance ...
Page 48: ...48 DSOX1204A G Oscilloscopes Service Guide 5 Replacing Assemblies ...
Page 50: ...50 DSOX1204A G Oscilloscopes Service Guide 6 Replaceable Parts ...
Page 56: ...56 DSOX1204A G Oscilloscopes Service Guide Index ...