Parallel and series connections (E36312A and E36313A only)
Parallel connections
Only connect outputs that have identical voltage and current ratings in parallel.
Connecting outputs in parallel provides a greater current capability than can be obtained from a single output.
The following figures show how to connect two outputs when in Auto-Parallel mode.
Front panel
Rear panel
The figure on the left illustrates 2-wire sensing. If voltage drop in the load leads is a concern, the figure on the right
shows how to connect the sense leads directly at the load.
30