Step 4: Install the Card

Diagram showing fan/support assembly fixings (U53xxA shown as example).

Installing the card retainer (if you have order your product with a card
retainer)

1. Slide the card retainer onto the fan card and fit screw (A) to the top of the retainer.

Removing the card retainer

1. Remove the screw (A in the figure).

2. Withdraw the card retainer.

Reversing the orientation of the fan assembly

It is also possible to reverse the orientation of the fan assembly, for instance in the case that
two cards must be placed in adjacent slots, or other PC hardware obstructs the fitting. This
procedure is detailed below.

1. Remove the two screws (B) from the cover maintaining the cables (see picture

above).

2. Remove the cover and lift up the cables to access the screws (C) below (see pic-

tures below).

3. Remove the four screws (B) nearest to the front of the fan unit.

4. Carefully, slide the fan unit approximately 15 mm to the rear until it clears the circuit

board, disconnect the power cables, and rotate the fan unit by 180°.

5. Reconnect the power cables.

6. Slide the fan unit back towards the card until it is fully seated into the support slots.

7. Refit the four screws (B), push the fan unit upwards away from the circuit board and

tighten the four screws (0.5 Nm torque).

8. Put back the cables and the cover.

9. Refit the four screws (C).

14

Acqiris U5309A Startup Guide

Summary of Contents for Acqiris U5309A

Page 1: ...Acqiris U5309A Acquisition Card 2 or 8 channels 8 bit 500 MS s to 2 GS s DC to 500 GHz bandwidth with real time processing Start up Guide ...

Page 2: ...ternational copyright laws Manual Part Number U5300 90016 Contact us Feel free to contact Acqiris experts and send an email to support acqiris com Worldwide contact for support or repair For any question of inquiry you can also contact your local office Acqiris Americas contact americas acqiris com Acqiris Europe contact emea acqiris com Acqiris Asia Pacific contact asia pacific acqiris com Acqiri...

Page 3: ...t is intended for use by qualified personnel who recognize shock hazards and are familiar with the safety precautions required to avoid possible injury Read and follow all installation operation and maintenance information carefully before using the product If this product is not used as specified the protection provided by the equipment could be impaired This product must be used in a normal cond...

Page 4: ...SI states that a shock hazard exists when voltage levels greater than 30V RMS 42 4V peak or 60VDC are present A good safety practice is to expect that hazardous voltage is present in any unknown circuit before measuring Operators of this product must be protected from electric shock at all times The responsible body must ensure that operators are prevented access and or insulated from every connec...

Page 5: ...trument or test fixture panels or switching card If you are using a test fixture keep the lid closed while power is applied to the device under test Safe operation requires the use of a lid interlock Cards and accessories shall not be connected to humans Before performing any maintenance disconnect the line cord and all test cables Any part or component replacement must be done by Acqiris No opera...

Page 6: ...ectronic product in domestic household waste Product Category With reference to the equipment types in the WEEE directive Annex 1 this product is classified as Monitoring and Control instrumentation product Do not dispose in domestic household waste To return unwanted products contact your local Acqiris office This symbol indicates the time period during which no hazardous or toxic substance eleme...

Page 7: ... Software 11 Step 4 Install the Card 11 Fitting the card with the optional card retainer into your PC 13 Reversing the orientation of the fan assembly 14 U5309A Front Panel Features 15 Front Panel Connectors 15 Front Panel LEDs 16 Step 5 Verify Operation of the U5309A Card 16 Keysight Connection Expert 16 Driver Graphical Interface MD3 Soft Front Panel 17 Perform a Verification of the U5309A optio...

Page 8: ...ns Electrostatic discharge ESD can damage or destroy electronic components Use a static safe work station to perform all work on electronic assemblies A static safe work station uses for instance two types of ESD protection conductive table mat and wrist strap combination and conductive floor mat and heel strap combination Both types when used together provide a significant level of ESD protection...

Page 9: ... labelislocatedonasticker Onthetopedgeofthecard for unitsshippedafter July2018 Onthesidepanelofthecard for unitsshippedbeforeJuly2018 Theserialnumber canalsobereadfromtheSoftFrontPanelinterface butonlyafter the hardwareandsoftwareareinstalled Nameandaddressofowner aP O boxisnotacceptableasareturnaddress Adescriptionofthefailureor servicerequested 4 Pack the card in its original ESD bag and packing...

Page 10: ... listed above is required for installation The NET Framework Runtime Components are installed by default with most Windows installations so you may not need this amount of available disk space Hardware Requirements A PC running one of the above operating systems Recommended models are HP Z420 HP Z440 or Dell T5810 Install the Software Keysight IO Libraries Suite IOLS The Keysight IO Libraries Suit...

Page 11: ...w keysight com main software jspx ckey 2364464 In this case please refer to the associated documentation available with MD2 installer Startup guide User s manual etc Acqiris MD3 Software for Signal Acquisition Cards includes device drivers IVI C IVI NET 1 and documentation 1 Download Acqiris MD3 from Acqiris Extranet 2 After downloading Acqiris MD3 Software for Signal Acquisition Cards double clic...

Page 12: ... adapters If your PC has spare 4 pin Molex power connectors use the supplied adapter 8121 2531 Note that this 12V power cable if for use with U5309A only If your PC only has the newer SATA power connectors use the supplied adapter 8121 2533 5 Replace any covers and switch on the PC Check the card front panel indicators a few seconds after the boot process both the Status and DPU LEDs should be gre...

Page 13: ...ional card retainer into your PC 1 Loosen but do not remove the screw A holding the rear card retainer and slide it towards the fans Note that there are screws only on one side there is no screw on the bottom side 2 Insert the card into the PCIe slot of the PC ensuring that it is fully seated into the PCIe bus connector 3 Fix the card to the PC either by fitting a screw or by using a method compli...

Page 14: ...he fitting This procedure is detailed below 1 Remove the two screws B from the cover maintaining the cables see picture above 2 Remove the cover and lift up the cables to access the screws C below see pic tures below 3 Remove the four screws B nearest to the front of the fan unit 4 Carefully slide the fan unit approximately 15 mm to the rear until it clears the circuit board disconnect the power c...

Page 15: ...SandTTLcompatible CLK IN MMCXfemale Externalclockinput AC coupledand50Ωterminated signallevel 5to 15dBm Pleaserefer todatasheetfor details REF IN MMCXfemale Externalreferenceclockinput AC coupledand50Ωterminated Itcanaccepta100MHzsignalfrom 3to 3 dBm Note It is recommendedtofirst connect theSSMC endof theinput cableandtightenusing thetorquewrench U1092A WCK beforeconnectingtheSMA end this minimize...

Page 16: ...ve the power sequence requirements If the error persists please contact technical support support acqiris com Step 5 Verify Operation of the U5309A Card Keysight Connection Expert The intention of this step is to verify correct operation of the newly installed card Run Keysight Connection Expert by clicking the task bar icon and select Connection Expert The list of installed cards is displayed The...

Page 17: ...ver Graphical Interface MD3 Soft Front Panel The AqMD3 IVI driver provides access to the functionality of AqMD3 ADC cards through a NET or ANSI C API which also complies with the IVI specifications The AcqirisMD3 SFP Soft Front Pannel is a graphical interface for signal acquisition carddrivers that enables the control of any supported ADC Cards The MD3 SFP can be launched from the Windows Start Me...

Page 18: ...l acquired from an external Function Generator Required Hardware An external signal source is required Almost any sine wave or function generator capable of generating a signal with an amplitude of 300 mV rms into 50 Ω at a frequency of 1 MHz may be used Hardware Description RF AnalogSignalGenerator e g KeysightN5181B SSMC cable 50ΩCoaxialcablewithSSMC f A1mSSMC toSMAcableisavailableas anaccessory...

Page 19: ...rify utility checks the version of the Control FPGA firmware If the version is not up to date the tool will automatically propose to update the firmware using the Firmware Update Utility Once the Control FPGA firmware has been updated successfully please power off your computer and restart it again for the update to take effect You may then proceed with theAqMD3Verify utility as described in this ...

Page 20: ...cqiris MD3 Documentation U5309A U5309A_Datasheet pdf or fromStartup Menu Acqiris MD3 Documentation U5309A U5309A_Datasheet Soft Front Panel help sys tem Providesinformationontheuseofthedriver SoftFrontPanel C Program Files Acqiris MD3 Documentation MD3_SFP_Help chm or fromStartup Menu Acqiris MD3 Documentation MD3_SFP_Help IVI Driver reference help sys tem ProvidesdetaileddocumentationoftheIVI COM...

Page 21: ...ments The AqMD3 IVI NET driver can be used in Visual C Visual C CLI or Visual Basic NET development environment Program examples Once that Acqiris MD3 software is installed program examples can be found in C Program Files1 IVI Foundation IVI Drivers AqMD3 Examples 1 Or your installation path ...

Page 22: ...This information is subject to change without notice Acqiris SA 2017 2018 Thursday July 12 2018 Switzerland U5300 90016 www acqiris com ...

Reviews: