50
U9361C/F/G/M RCal Receiver Calibrator Measurement Guide
Demonstrations
9.
Disable the RF output on the signal generator.
RF
Output
>
Off
10.
Disconnect the cable from the signal generator and connect it to the RCal module.
11.
Navigate to the RCal Calibration Configuration Window.
Input Output
>
Calibration
>
Configuration
The Calibration Configuration Table should be unpopulated.