U9361C/F/G/M RCal Receiver Calibrator Measurement Guide
33
Demonstrations
This should generate a warning that the current instrument state does not match the applied
calibration.
To handle both paths, an additional path must be added to the Calibration Configuration table.
20.
Disable the RF output on the signal generator through
Meas Setup
>
Source
>
RF
Output
>
Off
21.
Disconnect the cable from the signal generator and connect it to RCal module.
22.
Navigate to the RCal Calibration Configuration window.
Input Output
>
Calibration
>
Configuration
23.
Duplicate the existing state and change the new state’s RF path. (You will need to scroll to
access some of these settings)
Select
Row 1
>
Duplicate Row
>
uW Path Ctrl
>
uW Presel Bypass
24.
Clear the Calibrate check box on row 1 and select Calibrate on the new row.