U9361C/F/G/M RCal Receiver Calibrator Measurement Guide
41
Demonstrations
Another state must be added in order to properly apply corrections to both RF paths.
15.
Disable the RF output on the signal generator.
RF
Output
>
Off
16.
Disconnect the cable from the signal generator and connect it to the RCal module.
17.
Navigate to the RCal Calibration Configuration window.
Input Output
>
Calibration
>
Configuration
18.
Duplicate the existing state and change the new state’s RF path.
Select
Row 1
>
Duplicate Row
>
µW Path Ctrl
>
Standard