U3042AM08
General Specifications
General Specifications
Specifications for the Test Set are characteristic for the system performance of the
analyzer and Test Set. Actual performance of the system is based on the customers
analyzer that is used with the Test Set. A functional certificate is only offered for the
Test Set.
When connected to a
PN
A
, th
is
Test Set will degrade the performance at the test
ports. The internal switch paths reduce test port power
and power to
the receivers.
This affects the test port power of the
network
analyzer and also reduces dynamic
range. The reflection tracking values measured
the analyzers dynamic range to determine the approximate performance of the
system.
Power Requirements
Verify that the required AC power is available before installing the Test Set to
the analyzer.
• 100/120/220/240 V (50/60Hz)
• The instruments can operate with mains supply voltage fluctuations up to ± 10%
of the nominal voltage.
• Air conditioning equipment (or other motor–operated equipment) should not be
placed on the same ac line that powers the Test Set and analyzer.
• U3042M08 maximum power is 350 W.
WARNING
This is a Safety Class I Product (provided with a protective earthing ground
incorporated in the power cord). The mains plug shall be only be inserted in a
socket outlet provided with a protective earth contact. Any interruption of the
protective conductor inside or outside of the instrument, is likely to make the
product dangerous. Intentional interruption is prohibited.
12
Keysight U3042AM08 User's and Service Guide