U3042AM08
Controlling the Test Set with N5222A/B, N5230C, N5232A/B or N5242A/B
Channel: (4)
Channel Control
Label: (5)
Before Sweep Start
– After Sweep End
Tabs: (6)
Dwell After
Command: (7)
Handler I/O Control
and Aux I/O Output
Voltage: (8)
Reset All:
Save and Recall:
OK:
Cancel:
Help:
Specifies the channel number for dialog settings. Each channel is configured
individually. The drop-down list illustrates the channels that currently have
measurements. There must be at least one displayed trace for the Test Set I/O
Interface to function.
Specifies the label to be displayed on the analyzer’s screen during the channel
sweep.
Commands (GPIB, I/Os and Dwell) can be sent Before Sweep Start and After Sweep
End. However, they are configured and enabled separately on the “Interface Control”
dialog box. For example; to send a command before and after an analyzer sweep, the
“Enable Interface Control” check box must be selected and commands entered in
both the Before Sweep Start and After Sweep End tabs. The
Before Sweep Start
data
is sent before the first trace on the channel begins sweeping. The
After Sweep Start
data is sent after the last trace on the channel sweep is completed.
Specifies a wait time, in milliseconds, after all commands to all interfaces are sent.
Any positive integer is allowed. This is used to allow all external devices to settle
before beginning a measurement. An erratic trace could indicate that more settling
time is necessary.
Provides I/O interface control through the rear panel of the analyzer. Refer to the
Help menu for further information.
Sets all fields on all
channels to their default values.
Saves and recalls the contents of the dialog box. If the “Interface Control” dialog box
is populated with settings during an Instrument State Save, the settings are
automatically recalled with the instrument state settings. Interface control uses an
*.xml file type. An example file is stored on the analyzer’s hard drive. You can recall it
into the dialog, or you can open and edit it with a word processor, such as Word Pad.
Applies the settings and closes the dialog box.
Does not apply changes that were made and closes the dialog box.
Provides additional information for using the interface control application.
Keysight U3042AM08 User's and Service Guide