Verification Limits
Specifications for the test set are typical. System performance for the network
analyzer and test set are only characteristic and intended as non-warranted
information. Only a functional certificate is provided for the test set.
It is recommended that you return your instrument to Keysight Technologies for
servicing or repair if the test set and network analyzer performance exceed the
operational verification limits.
A periodic calibration is not required. The Operators Check should be performed after
System Setup, or if performance is in question. An N-Port calibration should be
performed before making a measurement.
Typical specifications are based on 1 to 2 units performance. Refer to
.
Table 1
7
PNA-L Reflection Tracking Limits
a
a. Reflection Tracking takes into account Source Loss, Receiver Loss, Margin,
and PNA Mixer Cal.
Frequency
Standard
Port 1- 16
b
b. A standard unit is one without Options 001 or 002
Option 001 or 002
Port 1-16
10 MHz to 4 GHz
-10 dB
0 dB
4 GHz to 6 GHz
-13 dB
-2.5 dB
6 GHz to 10 GHz
-16 dB
-5.0 dB
10 GHz to 15 GHz
-17 dB
-7.5 dB
15 GHz to 20 GHz
-25 dB
-15 dB
Table 1
8
PNA or PNA-X Reflection Tracking Limits
a
a. Reflection Tracking takes into account Source Loss, Receiver Loss, Margin,
and PNA Mixer Cal.
Frequency
Standard
Port 1- 16
b
b. A standard unit is one without Options 001, 002.
Option 001 or 002
Port 1-16
10 MHz to 4 GHz
-10 dB
0 dB
4 GHz to 6 GHz
-13 dB
-2.5 dB
6 GHz to 10 GHz
-16 dB
-5.0 dB
10 GHz to 15 GHz
-17 dB
-7.5 dB
15 GHz to 20 GHz
-25 dB
-15 dB
20 GHz to 26.5 GHz
-30 dB
-40 dB
U3042AE12 User's and Service Guide
61