Cal Kit Operational Check
The following procedure can be used to confirm that the test set and network
analyzer are operational. The operation verification limits provided ensure that your
test set and network analyzer is operating properly.
Equipment Required
The U3042AE12 requires that the user be familiar with the equipment and
components listed in
This section provides an equipment list and setup of the network analyzer and test
set.
Table 1
6
Equipment List
Description
Qty
N4691A 3.5 mm ECal Module 10 MHz to 26.5 GHz (Option 00F or M0F)
or
N4691B 3.5 mm ECal Module 300 kHz to 26.5 GHz (Option 00F or M0F)
or
Mechanical cal kit 85052B
or
85052D
1
N5230C 4-Port Network Analyzer (Option 245 and 551)
or
N5222A/B, N5232A/B
or
N5242A/B Option 400 and 551
1
Set of interconnect cables (PNA and test set), see
“PNA and PNA-X RF Interface Cable
1
60
U3042AE12 User's and Service Guide