2.
On the analyzer select
Response
>
Cal Wizard
-OR-
(Response > Cal > Other Cals...)
.
a.
If using a mechanical cal kit, select
SmartCal (Guided Calibration)
>
Next
.
b.
If using an ECal module, connect the ECal to an available USB port and select
Use Electronic Calibration (ECal)
>
Next
-OR-
(Response > Cal > Other Cals > ECal...)
.
Figure 3
5
Begin Calibration
3. Continue following the Cal Wizard prompts. In the “Select Calibration Ports and
ECal Module” dialog box press
Select All
, or select the ports you are calibrating
and
press
Next
. If an ECal characterization has been done, select
View/Select
ECal Module
and select the previously saved user file from
Figure 3
6
8-Port Calibration
4.
Connect the ECal or the mechanical cal kit to the ports you are calibrating
following the Cal Wizard prompts and click
Measure
after each connection. The
electrical delay value may be shown in the dialog box after the last measurement,
click
OK
.
5.
At the Calibration Completed prompt, select
Save As User Calset
and
type the
name desired and
Save
. See
.
6.
After calibrating Test Set ports, use a quality load and short to verify the
calibration on each port or end of the test cable. Measure reflection and confirm
the return loss is as expected. If the result is not as expected, repeat the
calibration without the Test Set and ensure that the network analyzer is in
standard (non-multiport) mode.
If measurement errors occur, ensure the newest version of firmware is installed on
the analyzer. Measurement errors can be a result of firmware algorithms. Consult
with Keysight Service or firmware web page for the latest firmware revisions and
history.
http://na.support.keysight.com/pna/firmware/firmware.html
U3042AE12 User's and Service Guide
43