Installation Note
N9000-90032
13
Option T03, T06 - Tracking Generator Installation Kit
3. Update the Instrument Software
1. Load the latest instrument software is required to assure all FPGAs and drivers located on both the
newly installed hardware and on the case instrument are synchronized. Therefore, even if the
instrument contains the latest revision of software, you must re-install the software to assure proper
operation.
The latest revision of software may be downloaded from:
http://www.agilent.com/find/Xseries_software
4. Install the License Files
1. Located Option Upgrade Entitlement Certificate in the kit and follow the directions to redeem it.
You will receive and email with an attached License File.
2. Located the USB Storage Device. Perform a virus scan on this device before use. Save the License
File to the root directory of the USB Storage Device.
3. Connect the USB Storage Device to the signal analyzer USB port.
4. The signal analyzer will automatically consume the License File. (This may take a few minutes)
When the License File is consumed, the Agilent License Manager will display a “Successful
License Installation” message as shown in
Figure 10
Successful License Installation
5. Tracking Generator Function Activation
After the installation is completed, please follow the procedures below to activate the option.
1. Press
[Source] > {Select Source}
.
2. Press arrow key to highlight
None
, and press
{Select Highlighted Source}
.
3. Press arrow key to highlight Agilent Tracking Generator, and press
{Selected Highlighted Source}
.