6
Installation Note
N9000-90032
Option T03, T06 - Tracking Generator Installation Kit
2. Hardware Installation
CAUTION
Electrostatic discharge (ESD) can damage or destroy electronic components. All
work on electronic assemblies should be performed at a static-safe workstation.
Refer to the documentation that pertains to your instrument for information about
static-safe workstations and ordering static-safe accessories.
Removal
1. Turn off the instrument and remove the power cord from the rear panel.
2. Remove the outer case of the instrument.
a. Refer to
. Using the T-20 driver, remove the four screws (two on each side) (1) that
attach the handle strap (2) on each side of the instrument.
b. Using the T-20 driver, remove the four screws (including washers) (3) that hold the rear feet (4)
in place.
c. Pull the instrument cover (5) off towards the rear of the instrument.
Figure 1
Standard Instrument Outer Case Removal
3. Remove the Top Brace and Reference Bracket.
a. Refer to
. To remove the top brace (1), use the T-10 driver, remove the eight screws (3)
(0515-0372) attaching the top brace to the chassis, the six screws (4) (0515-1227) attaching the
top brace to the boards, and the four screws (4) attaching the top brace to the reference bracket.