Keysight N778xC Series Polarization Instruments Programming Guide
73
Polarization Commands
5
Command:
:POLarimeter:SWEep:TRIGger:POST:SAMPles
Syntax:
:POLarimeter:SWEep:TRIGger:POST:SAMPles
Description:
Sets the number of post samples. That means samples after trigger event.
Numbers of pre samples and post samples have to be smaller than 1048576.
Parameters:
Number of post samples
Example:
:pol:swe:trig:post:samp 2000
Affects
N7781C / N7786C / N7788C
Command:
:POLarimeter:SWEep:TRIGger:POST:SAMPles?
Syntax:
:POLarimeter:SWEep:TRIGger:POST:SAMPles?
Description:
Returns the number of post samples
Parameters:
None
Response:
Number of post samples
Example:
:pol:swe:trig:post:samp?
→
+2000
Affects
N7781C / N7786C / N7788C
Command:
:POLarimeter:SWEep:CONTinuous:SACCuracy
Syntax:
:POLarimeter:SWEep:CONTinuous:SACCuracy
Description:
Sets the SOP accuracy in rad used in the continuous sampling mode.
A value of 0 causes all samples to be stored.
Otherwise SOP changes smaller than SOP accuracy will not cause a sample to be stored.
Parameters:
SOP accuracy in rad
Example:
:pol:swe:cont:sacc 0.1
Affects
N7781C / N7786C / N7788C