Keysight N778xC Series Polarization Instruments Programming Guide
71
Polarization Commands
5
Command:
:POLarimeter:SWEep:SRATe?
Syntax:
:POLarimeter:SWEep:SRATe?[<wsp>MIN|MAX|DEF]
Description:
This command gets the sampling rate and averaging time
Parameters:
None: to get current sampling rate and averaging time
MIN: minimum sampling rate
MAX: maximum sampling rate
DEF: default sampling rate
Response:
No parameter:
MIN|MAX|DEF parameter:
sampling rate in float
averaging time in float
sampling rate in float
Example:
:pol:swe:srat?
→
+1e+06,+1e-06
Affects
N7781C / N7786C / N7788C
Command:
:POLarimeter:SWEep:LPRange?
Syntax:
:POLarimeter:SWEep:LPRange?
Description:
Returns how well the ADC range was used for the last peak power. If this value is below 0.5, the measurement sweep should
be repeated with a higher amplifier gain setting. If the Value is > 1.0 then there is an increasing Risk of Overflow.
Select a lower amplifier setting in this case.
NOTE: use only gain between 0 and 7 for Stabilizer Mode because 8 and 9 have a
small Bandwith
Parameters:
None
Response:
Float
Example:
:pol:swe:lpr?
→
+0.5000
Affects
N7781C / N7786C / N7788C
Command:
:POLarimeter:SWEep:STEP
Syntax:
:POLarimeter:SWEep:STEP<wsp><value>[m|nm|pm]
Description:
Sets the sweep step in m