16
Keysight N778xC Series Polarization Instruments Programming Guide
1
Introduction to Programming
Common Status Information
There are three registers for the status information. Two of these are
status-registers and one is an enable-registers. These registers conform to
the IEEE Standard 488.2-1987. You can find further descriptions of these
registers under
*ESE
,
*ESR?
, and
*STB?
.
shows how the Standard Event Status Enable Mask (SESEM) and
the Standard Event Status Register (SESR) determine the Event Status Bit
(ESB) of the Status Byte.
Figure 1
The Event Status Bit
The SESR contains the information about events that are not slot specific.
0
1
2
3
4
5
6
7
*STB? returns the Status Byte Register
Status
OSB ESB
QSB
*ESR? returns the Standard Event Status Register
0
0
1
0
1
2
3
4
5
6
7
Event
1
0
0
0
0
0
Status
Register
0
1
2
3
4
5
6
7
Event
1
1
1
1
1
1
Status
Enable
Mask
*ESE sets the Standard Event Status Enable Mask
&
&
&
&
&
&
&
&
OR
Byte
All bits shown as
are unused
0
MAV