4-44 Keysight N5224A/25A Service Guide
Troubleshooting
Measurement System Troubleshooting
4-
—
“Bottom RF Cables, Standard 4-Port Configuration, Option 400” on
—
“4-Port Configuration, Option 419” on page 6-84
—
“Bottom RF Cables, 4-Port, Option 419 (Ports 1 and 2)” on page 6-86
—
“Bottom RF Cables, 4-Port, Option 419 (Ports 3 and 4)” on page 6-89
Getting Ready to Test
Before checking the assemblies, you must open the analyzer.
1.
Turn off the analyzer power.
2.
Unplug the power to the analyzer and disconnect all front and rear panel
connections except installed jumpers.
3.
Remove the outer and inner covers from the analyzer. Refer to
.
4.
With the covers off, plug in the analyzer and turn on the power.
Checking the A27 and A28 Mixer Brick Receiver Outputs
Set the network analyzer for an S
11
measurement with a CW frequency of 1
GHz.
1.
Locate the following flexible cables at the receiver IF outputs of the mixer
brick(s).
—
2-port models
—
Receiver A; W61 of A27
—
Receiver R1; W62 of A27
—
Receiver R2; W63 of A27
—
Receiver B; W64 of A27
—
4-port models
—
Receiver A; W61 of A27
Use an antistatic work surface and wrist strap to reduce the chance of
electrostatic discharge for all of the procedures in this chapter.
Procedures described in this document are performed with power
supplied to the product while protective covers are removed. Energy
available at many points may, if contacted, result in personal injury.
Do not operate the analyzer with the outer cover removed for more than
30 minutes, as this could cause the analyzer to overheat which could
result in costly damage.