Installation Note N5222-90114 65
Installation Procedure for the Upgrade
Connect New N5240-60098 (x1) Port 3 and Verify N5240-60098 Port 1
Blue Bias-Tee Combiner Cables and Clamps from the A71 and A72 Bias
Tees “RF-IN” to the A70 LFE Board “Port1” and “Port3” Connectors and
Install Clamps onto the Ferrite Beads
4.
Install the N5240-60098 (x2) cable as shown. Note the orientation of the
cable (item
through
).
5.
Route cables as shown.
Figure 38
Connect new N5240-60098 (x1) from Port 3 and verify N5240-60098 from
Port 1 Bias Tee Combiners to A70 LFE Board and install Clamps onto the
Ferrite Beads (N5240-60098 (x2), and 1400-1391 (x1)).
Verify Port
1 cable is
properly
routed and
secured.