![Keysight Technologies N5221A Configuration Manual Download Page 6](http://html.mh-extra.com/html/keysight-technologies/n5221a/n5221a_configuration-manual_1962033006.webp)
Page 6
Find us at
www.keysight.com
PNA Family Network Analyzer Configurations
(continued)
1 For N522xA and N5241/42/49A, vector-noise-corrected measurements require an ECal for use as an impedance tuner. For N5244/45/47A with Option 029,
an internal tuner is included. For calibration, Option 028 requires a power meter, and Option 029 requires either a power meter or a 346-series noise source
(Keysight 346C or 346C-K01 recommended). All options require a power meter for measuring mixers and converters.
2 Option 080 is required to configure an external source using External Device Configuration dialog.
3 Option 082 is a subset of Option 083; therefore, they cannot be ordered together.
4 A configurable test set is required for VMC measurements (to connect a reference mixer) or for SMC+Phase measurements using the comb-generator-based
calibration. When ordered with N522xA-200, -210, -400 and -410, Option 083 adds phase and delay measurements only by using SMC+Phase with a
calibration mixer.
5 Option 087 can be ordered without N524xA-224 or N524xA-423, but may require external equipment such as a signal generator and a combiner. Refer to
page 13 for more details.
6 When ordering a test set, select an appropriate interface kit. Refer to page 22 Multiport Measurements section for more details.
7 When configuring N524xA as a multiport analyzer using Option 551 and a multiport test set, the combiner feature of Option 224 or 423 is temporarily
disabled.
When configuring N524xA as a standalone analyzer, the combiner feature is enabled.
8 The Windows 7 upgrade kit for PNAs containing CPU version 6 is model N8983A. This upgrade kit contains a new solid state drive with all of the necessary
software installed. For units with older CPUs, order the N8984A with Option PC7. This upgrade kit contains a new solid state drive and a version 7 CPU. For
adding Option 090 to units with DSP version 4, add Option SP5 to the N8983A or N8984A order to get the DSP version 5.
9 Option 093 is a subset of 094; therefore they cannot be ordered together.
Application options
To add options to a product, order the corresponding item number (N52xxA-xxx).
Measurement application
For PNA-X
Series
For PNA
Series
For PNA-L
Series
Additional Information
Automatic fixture removal
N524xA-007 N522xA-007 N523xA-007 Requires Windows 7 OS
8
Time-domain measurements
N524xA-010 N522xA-010 N523xA-010
Dynamic uncertainty for S-parameter
measurements
N524xA-015 N522xA-015 N523xA-015 Displays the measurement uncertainty dynamically (real-time) on the
measurement trace. Not available with N5231A,32A and 39A
Noise figure measurements using standard
receivers
1
N524xA-028 N522xA-028
n/a
Requires Option 080. For measuring frequency converters, requires Option
082 or 083. Not available with N522xA-200, -210, -400 and -410
Fully-corrected noise figure
measurements
1
N524xA-029
n/a
n/a
Requires Option 080 and for N5241/42/49A, one of Options 219, 224, 419,
423 or H85. For N5244/45/47A, requires Option 224 or 423. On N5247A, noise
receivers work up to 50 GHz only. For measuring frequency converters, requires
Option 082 or 083
Frequency offset
2
N524xA-080 N522xA-080 N523xA-080
Scalar-calibrated converter measurements
3
N524xA-082 N522xA-082 N523xA-082 Requires Option 080
Vector- and scalar-calibrated converter
measurements
3, 4
N524xA-083 N522xA-083
n/a
Requires Option 080
Embedded LO measurements
N524xA-084 N522xA-084
n/a
Requires Option 082 or Option 083. Also works with Options 028, 029, 086
and 087
Gain compression application
N524xA-086 N522xA-086
n/a
For measuring frequency converters, requires Option 082 or 083
Intermodulation distortion
application
5
N524xA-087 N522xA-087
n/a
Requires Option 080 For measuring frequency converters, requires Option
082 or 083. Not available with N522xA-200, -210, -400 and -410
Source phase control
N524xA-088 N522xA-088
n/a
Not available with N522xA-200 and -210
Differential and I/Q devices application
N524xA-089 N522xA-089
n/a
Requires Option 080 and 400, 401, 410, 417, or 419, and Windows 7 OS
8
Spectrum analyzer
N524xA-090 N522xA-090 N523xA-090 Requires Option 080, Windows 7 OS, and DSP version 5
8
. Test set and power
configuration option with receiver attenuators is recommended
Extended spectrum analyzer to 110 GHz
9
N524xA-093 N522xA-093
n/a
Requires Option 090. Not available with N5241/49A or N5221A
Extended spectrum analyzer above 110 GHz
9
N524xA-094 N522xA-094
n/a
Requires Option 090. Not available with N5241/49A or N5221A
Integrated true-mode stimulus application
N524xA-460 N522xA-460
n/a
Requires one of Options 400, 401, 410, 417, 419 or 423
N-port capabilities
6
N524xA-551
7
N522xA-551
N523xA-551 Not available with N522xA-200, -210, -400 and -410, and N523xA-200 and -400
PNA-X Series
N5249A 10 MHz to 8.5 GHz
N5241A 10 MHz to 13.5 GHz
N5242A 10 MHz to 26.5 GHz
N5244A 10 MHz to 43.5 GHz
N5245A 10 MHz to 50 GHz
N5247A 10 MHz to 67 GHz
PNA Series
N5221A 10 MHz to 13.5 GHz
N5222A 10 MHz to 26.5 GHz
N5224A 10 MHz to 43.5 GHz
N5225A 10 MHz to 50 GHz
N5227A 10 MHz to 67 GHz
PNA-L Series
N5239A 300 kHz to 8.5 GHz
N5231A 300 kHz to 13.5 GHz
N5232A 300 kHz to 20 GHz
N5234A 10 MHz to 43.5 GHz
N5235A 10 MHz to 50 GHz