4
Operating Guide
38
Keysight N4000/1/2A Operating and Service Manual
3
Connect the SNS to the NFA Port using the multi-pin 11730 cable, as shown in
.
Figure 4-9
SNS connected to NFA graphic
4
Press the
Frequency/Points
key.
5
Press the
Frequency Mode
menu key.
6
Press the
Fixed
menu key.
7
Press the
Fixed Frequency
menu key. Enter 1.0 GHz as a fixed frequency value,
by pressing the
1
numeric key followed by pressing the
GHz
menu key.
8
Press the
Sweep
key.
9
Press the
Manual Meas
menu key.
10
Press the
Manual State
menu key and set it to
Manual State (On)
.
11
Press the
Noise Source
menu key and set it to
Noise Source (On)
.
12
Press the
IF Att
menu key and set it to
IF Att (Hold)
.
13
Monitor the power value which appears in the manual measurement screen.
shows a typical example of this.