Low-Frequency Compensation
N2840/1/2/3A and N2853A User’s Guide
17
4
If you are using a 2000X, 3000X, or 4000X InfiniiVision oscilloscopes, use the
guided probe compensation:
a
Press the
Probe
softkey.
b
Press the
Probe Comp
softkey to run the compensation.
5
If you are using a different oscilloscope, display two or three waveform cycles
over two to six vertical divisions.
6
Use the supplied trimmer tool to adjust the
LF Comp
to an optimum square wave
Figure 13
Proper Low-Frequency Compensation Adjustment
NOTE
If the probe is already properly adjusted, the guided
compensation will not prompt you to make the adjustment.