Do you have everything? 

 

FieldFox –To check for installed options, press 

System

 then 

System Configuration

 then 

Options

 then 

Show Options

  

 

Phase stable jumper cable or adapters – Used to 
connect the FieldFox to devices with various 
connector types and gender.  

 

Attenuators – Reduces power and prevents 
damage to the FieldFox at the RF IN port. 

 

Cal Kit – OSL Tee or other calibration devices. 

 

SD card or USB flash drive – Save files to these 
devices for quick download to a PC. 

 

USB Power Sensor – For making Power Meter 
measurements. (Opt 302) 

 

Battery charger or extra battery 

 

The Power Button and LED 

 

 

To turn power ON, briefly press the power button. 
The LED is green. The boot-up takes about 60 
seconds.  

 

To switch to Stand by power (low battery drain) 
briefly press the power button. The LED is amber and 
blinks slowly. To turn power ON, briefly press the 
power button.  Power and settings are restored 
instantly. 

 

To turn Power OFF (very low battery drain) press and 
hold the power button until power is shut OFF - 
about 4 seconds. The LED is OFF. 

 

 

 

Summary of Contents for FieldFox N9927-90002

Page 1: ...n and LED 1 Battery Usage 2 Measure Return Loss CAT 3 Measure 1 Port Cable Loss CAT 4 Measure DTF CAT 5 Measure S Parameters NA 6 Multi Trace Configurations NA 8 Calibration CAT NA VVM 9 Spectrum Anal...

Page 2: ...Save files to these devices for quick download to a PC USB Power Sensor For making Power Meter measurements Opt 302 Battery charger or extra battery The Power Button and LED To turn power ON briefly...

Page 3: ...he FieldFox display as much as possible Briefly press the power button to switch to Stand by mode when not being used Press again to restore power All current settings are preserved The FieldFox will...

Page 4: ...UT 2 Select Preset then Preset Returns the FieldFox to known settings 3 Select Mode then CAT Cable and Antenna Test 4 Then Return Loss Default measurement 5 Press Freq Dist and enter Start and Stop fr...

Page 5: ...ss Sweep 3 then Min Swp Time Increase the sweep time until a stable trace is visible on the screen 6 Press Cal 5 then follow the calibration prompts 7 Connect the DUT cable to be tested to the FieldFo...

Page 6: ...to Fault 5 Press Freq Dist then Stop Distance and enter the length of the DUT 6 If a diplexer or other filter is in the measurement path press Meas Setup then select Frequency Mode Bandpass Then pres...

Page 7: ...surement Requires Opt 122 OR Press Trace 6 then Num of Traces then select a Multi trace configuration Learn more on page 8 3 Press Freq Dist then either Start and Stop or Center and Span to enter a fr...

Page 8: ...cale Manually enter a scale per division to view specific areas of the trace Ref Level Manually set the value of the reference line Ref Position Manually set the position of the reference line Values...

Page 9: ...such as format markers and limit lines are applied to the ACTIVE trace in the same manner as when a single trace is present To select a multi trace configuration Press Trace 6 then Num Traces Then ch...

Page 10: ...or Power Level is changed When using a phase stable jumper cable or adapter to connect the DUT to the FieldFox These should be high quality components Press Cal 5 then follow the prompts During a Quic...

Page 11: ...increase with more attenuation then compression DID exist Set the RF Attenuation value at the setting when further increases no longer result in an increase in the displayed power level Set RF Attenu...

Page 12: ...hich allows you to use the arrows to increment the channel number by the specified value 5 Then Back and Center Start or Stop Channel 6 Press Measure 1 then Channel Measurements and choose from the fo...

Page 13: ...e choose where in the FieldFox data flow that data is recorded RawMeas is easiest Record Interval sets the amount of time to wait between individual trace recordings Segment Count sets the number of t...

Page 14: ...ed from the measurement using Offset as follows Select Offset Press Scale Amptd then Offset On Off Then Offset Val Enter a value by which the power meter display is offset A positive value compensates...

Page 15: ...d USB Saves Recalls files to or from a USB Flash drive 2 Then File Type State Saves ALL FieldFox instrument settings and calibration for ALL Modes to an sta file Trace State Saves the current trace al...

Page 16: ...Manufacturing Part Number N9927 90002 Print Date October 1 2014 Supersedes June 12 2012 Keysight Technologies Inc...

Reviews: